IR-564/301 Blackbody SystemThe perfect blend of innovative design and state of the art technology

IR-564/301 Blackbody System


The IR-564 extends the temperature range of the IR-563 to 1200 º C by changing cavity materials to Silicon Carbide and high purity Alumina ceramics, otherwise the two units are virtually identical.

The 20º tapered - recessed - cone, surface emissivity, and cavity apsect ratio combine to provide blackbody radiation by multiple reflection, absorption and re-emission of it's thermal energy. The thermal energy of the cavity is provided by a ceramic-sealed heater coil that uniformly heats the cavity cylinder to temperatures from 50 º C to 1200º C. The IR-564 system has a full, one-year warranty.


  SYSTEM SPECS

 FEATURE  SPECIFICATION
Temperature Range

50 – 1200 C

Emittance Watts/Cm^2 (Watts)

26.68  (135)

Wavelength Range

0.5 - 99um

Emitter Size: in (mm)

1" (25.4)

Source Type

Cavity

Temperature Resolution

0.1 C

Calibration Accuracy

+/- 0.2 C to NIST Standard

Stability: Short (Long) Term

+/- 0.1 C (+/- 0.2C)

 

Response Time

> 100-1200 <70 Minutes

Temperature Sensors

Embedded 0.01% Matched Type S

Control Type

Active Multi-Band P.I.D.

Line Voltage

90 to 125 or 208-240 VAC 50-60 Hz

Power Requirements

800 Watts Max

Cable Length

8 Feet (2.4 m)

Dimensions: in (mm) Source

11.75” H x 11.4” D x 8” W (298x289x203)

Controller

5.1"H x 13.4"D x 12"W (130x340x304)

Warranty

1 Year

Standard Apertures

0.0125, 0.025, 0.050, 0.10, 0.20, 0.40, 0.60 and 1.0 In

Remote Interface

RS-232 or IEEE-488/GPIB

Weight

Source: 17lbs Controller: 9 lbs (Rackmounted 10 lbs)

Shipping Weight

2 Boxes; 35 lbs